• Use only hospital-approved disinfectants to clean the adapter. The
following solutions are acceptable for cleaning or disinfecting the adapter:
water, Betadine™, hydrogen peroxide, and mild detergent. After cleaning,
wipe the adapter with a sponge or soft cloth and water.
• Immediately remove any contamination from the adapter by using a
sponge or soft cloth and water.
• If contamination cannot be removed with a sponge or soft cloth and water,
the adapter may be immersed in water for up to 10 minutes. Do not
immerse the adapter in any other liquids.
10 Service
Warning: The adapter is not serviceable. Do not disassemble the adapter
because the performance of the adapter may be adversely affected. If the
adapter has been tampered with, do not use it.
11 Specifications for the Affinity CP adapter
Part number:
Compatible pump:
Compatible sys-
tems:
Maximum adapter
speed:
IP rating:
11.1 Gross dimensions
Weight:
Height:
Diameter:
AP40AST
• Affinity CP centrifugal blood pump Model AP40
• Affinity CP centrifugal blood pump Model BBAP40
• Affinity CP centrifugal blood pump Model CBAP40
• Stöckert and Sorin centrifugal pump systems (with
drive unit Model 60-01-04 and Model 60-01-00
and emergency drive unit Model 60-01-35 and
Model 60-01-50)
• Sarns and Terumo centrifugal systems (with drive
unit Model 164267 and manual drive unit
Model 164268)
3600 rpm
IP47
315 g (0.69 lb)
28 mm (1.1 in)
94 mm (3.7 in)