6 – GUI DE FO R ACCU R ATE ME AS U RE ME NT
Turn on the Apex ID prior to use.
Plug the Checker [Fig 1-9] (provided in the original packaging)
into the probe cord jack [Fig 2-5] and confirm the value on
the LCD screen is 0.5.
It is recommended to insert the file into the canal first and
then connect the file to the file holder.
If the signals do not change as the file approaches the
APEX, it is possible the root canal is too dry; therefore it
should be filled with saline.
Necrotic tissue or other matter in the root canal may affect
accuracy. It is advisable to take several measurements
during enlargement of the canal.
Start with a larger file at first. If the APEX indication (graph
and numeric signal on the screen) does not reach the
APEX, gradually try a smaller file to measure the working
length of the root canal.
7 – MAIN TEN AN CE A N D STO RAG E
7.1 – Sterilization
The 'Lip Hook' and 'File holder A' should be sterilized at
121°C for 30 minutes with high-pressure steam.
The 'File holder B' should be sterilized at a temperature
of 121°C for 30 minutes with high-pressure steam in a
sterilization pouch, then dried at least 10 minutes after
sterilization.
12
CAUTION Parts used in direct intraoral
contact are to be steam sterilized.
Type B
Caution
This
Stack 12
Dry
Applied
For Dental
Side Up
Maximum
Part
Use Only
12
NOTICE It is highly recommended to place
File holder B in a sterilization pouch for autoclaving.
Type B
Caution
This
Stack 12
Dry
Applied
For Dental
Side Up
Maximum
Part
• The metal of the File holder B may be corroded
Use Only
when it is in direct contact with steam.
• Temperature should not exceed 135°C during
sterilization as the product may be damaged.
7.2 – Recommended sterilizer type
• Sterilizer: Vacuum type (B)
• Sterilization temperature: 121°C,
• Sterilization time: 30min
9
Rx ONLY
Rx ONLY
Temperature
Consult
Dispose
0120
Instructions
-20C to 50C
Humidity
For Use
Properly
5% to 90%
To minimize error, it is recommended the working length
be measured more than twice.
If root canals within a tooth are connected, as in the case
of multiple root canals with anastomoses or cracks, miss-
measurement is possible. It is recommended to verify the
tooth's anatomy radiographically.
Contact with metal restorations by the file during
measurements may result in errors.
Remove blood or exudate from the canal before
measurement.
In cases where the foramen is not completely formed or
the foramen has been overly expanded, the measurement
value of the working length may be shorter than the actual
measurement value.
7.3 – Method used to validate recommended
sterilization method:
Sterility Assurance Level (SAL): 10
1) Sterilization validation samples are wrapped with BIs.
2) 3 times of sterilization cycles as a half cycle are carried
out at 121°C for 15 minutes.
3) 1 time of sterilization cycle as a full cycle is carried out
at 121°C for 30 minutes.
4) Incubate BIs at 55~60°C for 7days.
7.4 – Sterilizer used for validation
• Manufacturer: DAIHAN LABTECH CO., LTD
• Model name: LAC-5105SP
• Serial No. : 20101202138
• Control No. : TE-10-23
• Valuable room: 100 liter
• Power supply: AC 220V, 50/60Hz
• Pressure: 2.6kgf/cm2
• Range of sterilization temperature: 105°C~135°C
7.5 – Cleaning
Clean the Apex ID or probe cord with dry gauze and a
small amount of ethanol. Any ethanol left on the product
after cleaning should be completely wiped off with a dry
cloth.
and Apex ID during operation, followed by removal
Handle
Keep
Type B
Caution
Of
With
Dry
Applied
For Dental
Care
Part
of the sleeve and wiping with alcohol after operation.
Use Only
12
CAUTION Use protective sleeves on the cords
Rx ONLY
This
Stack 12
Side Up
Maximum
-6

Publicité

Table des Matières
loading

Table des Matières