Cover replacement, top (Continued)
8.
Unzip the support surface top cover from the bottom using the zipper pull tool or equivalent (A)
page
1-21).
Note: The zipper starts on the patient right side near the foot end.
9.
Remove and discard the top cover.
10. Reverse to install the new cover.
Notes
After installation, remove the zipper pull tool or equivalent from the zipper.
Connect the corner retainers by placing the strap below each D-ring and snaps to the cover. Repeat for the
other three corners.
Make sure that the watershed covers the zipper.
11. Verify proper operation before returning the product to service.