Disinfecting - Stryker 2971 Isolibrium Manuel D'utilisation

Surface de soutien. destinée à une utilisation avec le châssis de lit intouch fl27 (2131, 2141, 2151, 2152, version logicielle 4.0 ou supérieure).
Masquer les pouces Voir aussi pour 2971 Isolibrium:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

English
WARNING
EN
Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.
Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning or
disinfecting. Some cleaning and disinfecting agents are corrosive in nature and may cause damage to the product if
you use them improperly. If you do not properly rinse and dry the product, a corrosive residue may be left on the
surface of the product that could cause premature degradation of critical components. Failure to follow these
cleaning or disinfecting instructions may void your warranty.
Do not allow fluid to pool on the support surface. Fluids may cause degradation of components and may cause
unpredictable safety and performance of this product.
Always unplug the product power cord before cleaning or disinfecting to avoid the risk of shock.
CAUTION
Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
Do not over expose the covers to higher concentration disinfectant solutions as these may degrade the covers.
Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the
cover or the CPR strap print.
Prerequisite: Minimum of two operators are required to disinfect the support surface.
Suggested disinfectants:
Quaternaries
Phenolic disinfectants
Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts water)
Chlorinated bleach solution for 2972 Endurance cover is 10,000ppm
70% Isopropyl Alcohol
To disinfect the support surface covers after each patient use, follow these steps in order:
1.
Unplug the InTouch power cord.
2.
Unplug the support surface.
3.
Disconnect the integration cable.
4.
Thoroughly clean and dry the support surface covers (see
5.
Apply recommended disinfectant solution with a damp cloth or pre-soaked wipes (do not soak the support surface).
Notes
Make sure that you follow the disinfectant manufacturer's instructions for appropriate contact time and rinsing
requirements.
To disinfect the bottom cover of the support surface, raise the head of bed and foot end to help fold the head
end and foot end of the support surface.
6.
Rinse the support surface covers with a clean, damp cloth to remove excess disinfectant.
7.
Wipe the support surface covers with a clean, dry cloth to remove any excess liquid or disinfectant.
8.
Allow the support surface covers to dry completely before returning to service.
1-18

Disinfecting

Cleaning on page
2971-209-001 REV B
1-17) before disinfectants are applied.
www.stryker.com