Calys60/80/120 series are two insulated channels, multifunction calibrators. They
are hand-held instruments developed to meet all the needs of instrumentation
engineers and Quality managers, both in laboratory and in fieldwork. These units are
accurate, rugged, compact and easy to use. They are the best solution to simulate
and measure electrical and physical parameters: Voltage, current, resistance,
thermocouple, resistance thermometers, pressure, frequency and pulse.
Advanced flexibility and high performance has been achieved using 32-bit
microprocessor and a fast A/D conversion technology. The calibrators memory, has
stored inside all data for normalised IEC, DIN and JIS thermoelectric sensors for both
IPTS68 and ITS90 International Temperature Scale. The microprocessor performs
automatic linearization and cold junction compensation to assure high accuracy. It is
possible to set the calibrator to execute menu-driven calibration procedure for your
instruments in fieldwork.
7 different models are available to meet different needs about test and calibration:
Modèle
CH1
CH2
Calys60 IS
IN
OUT
Calys 80 IS
IN
OUT
Calys 120 IS
IN
IN/OUT
IN /OUT
Calys 80 P IS
-
(V,mA,PT100)
Caly 60 P
-
IN /OUT
Calys 80 P
-
(V,mA,PT100
NA:Not applicable.
Both Channel 1 (IN) and Channel 2 (IN/OUT) have the following operative mode
capability: millivolts, volts, milliamperes (active and passive loop), ohms,
temperature with thermocouples, temperature with resistance thermometers,
frequency, and pulse (except for the calys 60 P and calys 80 P).
Optional single or dual sensors Pressure inputs can be installed inside the calibrator
for gauge, absolute and differential measurements.
Manuel d'intructions NTA47248-300A4
Internal
External
Hart Module
pressure
pressure
NON
OPTION
OPTION
OPTION
OPTION
OPTION
OPTION
OPTION
-
OPTION
OPTION
OPTION
OPTION
OPTION
OPTION
An external pressure sensor can be installed for measurements up to 700 bar.
Optional auxiliary measurements (EC module) with integrated sensors are: Relative
humidity and temperature, Barometric pressure sensor.
Report of Calibration
Each unit comes standard with a calibration certificate that lists the nominal and real
values and deviation errors.
EMC Conformity
The instrument fulfils the prevision of the directive 89/336/CEE Electromagnetic
Compatibility.
2.1

Accessories

EXTERNAL PRESSURE MODULES - AISI 316SS - ±0.025 % F.S.
GAUGE
EPM000100G
EPM000500G
EPM001000G
EPM002000G
NA
EPM007000G
EPM200000G
NA
EPM035000G
NA
EPM070000G
EPM150000G
NA
EPM700000G
ABSOLUTE
EPM002000A
EPM020000A
HAND PUMPS
H801
External pneumatic hand pump 0-5bar
LPT1
External pneumatic hand pump -0,85- 3 bar
TP140
External pneumatic hand pump –0.95 to 40 bar
HTP1 700
External Hydraulic hand pump 700 bar
61
from -100 to 100 mbar (1.5 PSI) res. 0.001 mbar
from -500 to 500 mbar (7 PSI) res. 0.01 mbar
from -0.95 to 1 bar (15 PSI) res. 0.01 mbar
from -0.95 to 2 bar (30 PSI) res. 0.01 mbar
from -0.95 to 7bar (100 PSI) res. 0.1 mbar
from -0.95 to 20 bar (300 PSI) res. 0.1 mbar
from -0.95 to 35 bar (500 PSI) res. 1 mbar
from 0 to 70 bar (1000 PSI) res. 1 mbar
from 0 to 150 bar (2000 PSI) res. 1 mbar
from 0 to 700 bar (10000 PSI) res. 10 mbar
from 0 to 2 bar (30 PSI) res. 0.01 mbar
from 0 to 700 bar (10000 PSI) res. 10 mbar