5.5.1
Frequency/Pulse: generation
•
Press [SELECT] until "Select CH2" appears
SEL ECT Ch 2
Ou t
I n
Of f
Pr e s s < > t o SET
Select "OUT"
Vo l t a g e
Re s i s t a n c e
Fr e q u e n c y / Pu l s e
Pr o c e s s Fu n c t i o n
Te mp Ge n e r a t o r
Pr e s s < > t o SET
Select Frequency/Pulse
Fr e q u e n c y / Pu l s e
Fr e q u e n c y
Pu l s e
Pr e s s < > t o SET
•
select "Frequency" or "Pulse"
•
Press the [ ] key to select the option. Press the [ESC] key to return in
measure mode without changes
If you select "Frequency" as output, the display shows:
Fr e q u e n c y
Sq u a r e
Si n u s o i d a l
Tr i a n g u l a r
Amp l i t u d e [ 5 ]
Si n o f f s e t
Pr e s s < > t o SET
Manuel d'intructions NTA47248-300A4
OUT
[ 0 ]
•
Select the Amplitude
•
Select the offset with "Sin offset"
•
Select the desired waveform: square, sinusoidal, triangular
•
Select the range for the frequency :
Pr e s s < > t o SET
•
Press the [ ] key to select the option and to enter measure mode
If you select "Pulse" as input, the display shows:
Pr e s s < > t o SET
•
Define the time (minutes and seconds), over that the generation is performed
with Time
•
Select the mode for pulse generation: "One shot" or "Continuous",
depending on which kind of generation you want: over a single time range or
continuously
•
Select the Amplitude
•
Select OnOff. The instrument comes back automatically at the measuring
page and channel 2 expresses the value in p/t. Hence it is possible to directly
set the number of pulses in time unit.
When the instrument is in the measures page, it appears in Hold mode. Unlocking
hold by pressing [HOLD], the generation starts with those previously set parameters.
5.6
X-Scaling setting
The calibrator can measure voltage or current and simply display it or convert it using
the "XScale" function and then display the measured value via a conversion factor.
•
Press the [MENU] key. The following page is shown:
89
Si n u s o i d a l
2 0 0 Hz
2 k Hz
2 0 k Hz
Pu l s e
On Of f
Ti me
Mo d e [ On e s h o t ]
Amp l i t u d e [ 5 ]