Switching On Settings (Power On) - AOIP calys 60 IS Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Scroll down and select "Alarm"
Co n f i g u r a t i o n
Da t e / t i me
Re j e c t i o n [ 5 0 Hz ]
Ba u d r a t e [ 9 6 0 0 ]
OC Te s t
Al a r m
Pr e s s < > t o SET
Ge n e r a l
Us e r
Pr e s s < > t o SET
Select General and then Enable or Disable the general system alarms
NOTE: when you disable the alarms, at next turning on the selection will be
lose and alarms become active.
See par.10.1 about error messages.
Select User to set alarms definable by the user:
On Of f
Ch a n n e l
Mi n [ 0 ]
Ma x [ 1 0 0 ]
Pr e s s < > t o SET
Set the channel where to have threshold alarms with Channel
Set Min and Max threshold
NOTE: unit measure is referred to the used channel
Enable the alarm with "OnOff"
NOTE: in case of threshold overcoming, an acoustic signal is emitted until the
signal comes back return under threshold or the alarm is disabled.
Manuel d'intructions NTA47248-300A4
[ En a b l e ]
Al a r m
[ En a b l e ]
Us e r
[ o f f ]
[ c h 1 ]

5.15 Switching on settings (Power On)

It is possible to set how the instrument turns on through the "Power On" function
Press [MENU]
Pr e s s < > t o SET
Select "Configuration"
Di s p l a y
Po we r On [ Se t
Da t e / t i me
Re j e c t i o n [ 5 0 Hz ]
Ba u d r a t e [ 9 6 0 0 ]
Pr e s s < > t o SET
Select "Power On"
>
Pr e s s < > t o SET
Safe: Ch1 = voltage, Ch2 = off. This way avoids any risk of damage for the
instrument
Set Ch: it set the last channel settings
Set val: it set the last channel settings and the channel 2 output value
97
MENU
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Co n f i g u r a t i o n
Ch ]
Po we r On
<
Sa f e
Se t
Ch
Se t
v a l