For a better understanding of the appropriate connection when using the instrument
to simulate current into industrial 2-wire loop please, note the meaning of the terminal
used.
Passive loop
This type of connection is to be used when the external loop is not equipped with the
loop power supplied. The calibrator can be, as an example, connected directly to a
recorder, controller, etc. with input circuits configured for current measurements.
Active loop
This type of connection must be used when the external loop is equipped with its
loop power supplied. The power supply is not required to be disconnected. The loop
circuit must be opened and the calibrator connections are placed in series on the
loop.
Manuel d'intructions NTA47248-300A4
+
-
74
3.4.1

Multi-connection binding post

Calys60/80/120 include 3 different connection systems:
Standard banana plugs
Mini isothermal TC's connector
Push & Lock system for wires
Isothermic connector
for TCs
Banana plug
Push the binding post
and insert the wire