Pr e s s u r e
Pl
Ph
Pe
Pr e s s < > t o SET
If you have to verify the switch on a temperature, select "Temperature" and choose
the proper channel.
Confirm and then perform the test, verifying whether the contact is Close or
Open
Swi t c h Te s t
1 0 . 0 mb a r
Co n t a c t
0 . 0
Swi t c h Te s t
0 . 0 mb a r
Co n t a c t
1 5 . 0
During a switch test, at the first line you can read the measure in real time. The
second line shows the status of the switch and the third line indicates the measure
held at the moment of switch commutation.

5.13 Leak Test

It is possible to perform a leak test of pressure (only in models with pressure
sensors).
Calys60/80/120 carries out 4 different pressure readings at settable interval times
(T1..T4). At the end, the difference between the last and the first readings, with the
aim of giving a leak measure.
Press [MENU], the following page is shown:
Manuel d'intructions NTA47248-300A4
Op e n
mb a r
Cl o s e
mb a r
Pr e s s < > t o SET
Select "Advanced"
Pr e s s < > t o SET
Select "Leak Test"
Pr e s s < > t o SET
OnOff: run/stop the test
Channel: set the pressure input for the test
Wait time: seconds before starting the leak test procedure
Leak Time: leak test interval (seconds)
Select "OnOff" to start the test
The set waiting time will be start and the following pages will appear
Pr es s ESC t o abor t
the test start and the set leak time will countdown;
95
MENU
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Ad v a n c e d
Swi t c h Te s t
L e a k Te s t
HART
Tr x s i mu l a t o r
Gr a p h
L e a k Te s t
On Of f
[ o f f ]
Ch a n n e l
[ Pe ]
W a i t
Ti me [ 1 0 ]
L e a k Ti me [ 6 0 ]
Leak Tes t
W ai t i ng Ti me 10
Ac t ual
= 100. 00