Calys 60 P, Calys 80 P - AOIP calys 60 IS Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Transmitter
Sources: Voltage, current, Temperature, Resistance
Response time: 0.1 / 1 / 10 sec selectable
Ramp / Cycle
Sources: Voltage, Current, Temperature, Resistance
Sampling time: MAX. 0.1 sec
DataLogger
Sources: Voltage, Current, Temperature, Resistance
Sampling time: MAX. 1 sec
Memory: >1500 readings complete with date and time
General
Accuracy: the above accuracies are stated for 365 days and includes non-linearity,
hysteresis, and repeatability. The average temperature coefficient, inside the
temperature compensated range is ±0.002 of rdg/°C (w.t.r. +23°C / +73°F).
Compensation temperature range: 0 to 45°C (+32 to +113°F)
Calibration: self learning technique with automatic procedure
Channel 1-Channel 2 insulation: 250 Vdc
Common mode rejection: 140 dB at ac operation
Normal mode rejection: 60 dB at 50/60 Hz
Temperature stability: for temperature exceeding the band +18°C to 28°C
Span: ±8 ppm/°C
Zero: ±0.2 µV/°C
Measurement sampling time: 250 ms
Display: graphic LCD display with automatic and manual backlight device
Digital interface: full bi-directional RS232
Internal data memory: standard 512 kb – Optional 8Mb with internal card
Power supply: external charger and rechargeable Ni-MH battery
Battery life (typical):
10 h (8 h on IS models) on Tc and mV input/output (backlight Off)
4 h (3 h on IS models) with 20 mA simulation (backlight Off)
Recharging time (typical): 5 h (8 h on IS models) at 90 % and 6 h (10 h on IS
models) at 99 % with instrument switched off.
Battery charge indication: bar graph on the LCD display (flashing on charge)
Line operation: 100V - 120 V - 230V - 240 Vac with the external battery charger
Line transformer insulation: 2500 Vac
Sealing: IP54
Operating environment temperature range: from -10 °C to +55 °C
Storage temperature range: from 0 °C to +60 °C (excluding batteries)
Humidity: max 95 %RH non condensing
Case: Injection moulded polycarbonate case
Weights: net 1.4 Kg
gross 2.5 Kg
Manuel d'intructions NTA47248-300A4
Dimensions: 290x98x57 mm
2.2.2
IN/OUT Voltage
RANGE
-20 to 200 mV
-0.2 to 2 V
-2 to 20 V
Input impedance:
>10 MΩ for ranges up to 2000 mV f.s.
>500 kΩ for ranges up to 20 V f.s.
Output impedance (emf output): less than 0.5 Ω with a maximum cu rrent of 0.5 mA
Output noise (at 300 Hz):
IN/OUT Current
RANGE
0.1 µA
-5 to 50 mA
Input impedance: <20 Ω at 1 mA
Limited to 21 mA in passive mode
Input impedance: <20 Ω at 1 mA
Channel source Max 24 V.
66

Calys 60 P, Calys 80 P

ACCURACY
RES.
Calys 80 P
1 µV
±(0.01 % rdg + 3 µV)
10 µV
±(0.01 % rdg + 10 µV)
100 µV
±(0.01 % rdg + 0.1 mV)
<2 µVpp for ranges up to 200 mV f.s.,
<10 µVpp for ranges up to 2000 mV f.s.
<80 µVpp for ranges up to 20 V f.s.
ACCURACY
RES.
Calys 80 P
±(0.01 % rdg + 0.4 µA)

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Calys 80 isCalys 120 isCalys 80 p isCalys 80 pCalys 60 pCalys 60 p is

Table des Matières