TC TYPE
RANGE
RESOL.
Tc L
-200 to 760 °C
0.1 °C
-330 to 1400 °F
0.1 °F
Tc N
-270 to 1300 °C
0.1 °C
-450 to 2380 °F
0.1 °F
Tc E
-270 to 1000 °C
0.1 °C
-450 to 1840 °F
0.1 °F
Tc F
0 to 1400 °C
0.1 °C
32 to 2560 °F
0.1 °F
Note: 0.1°C resolution with temperature lower than –200°C
Selection °C/°F/K: through the configuration procedure
Resolution: 0.01°C / 0.01°F
Temperature scale: ITS90 and IPTS68 selectable
Reference junction compensation:
internal automatic from -10 °C to +55 °C
external adjustable from -50 °C to +100 °C
remote with external Pt100 from -10°C to +100 °C (only on XP model)
Rj compensation drift: ± 0.002°C/°C (from -10 °C to +45 °C)
Rj accuracy: ±0.05°C @ 25°C / ±0.005 °C/°C
Input impedance (Tc ranges): >10 MΩ
Pressure (option)
Pressure fluids: AISI 316 SS compatible fluids (water, gas, and oil)
Temperature compensation: Automatic with built-in calibration matrix from 0°C to
50°C.
Engineering units: mbar, bar, Pa, hPa, kPa, MPa, kg/cm2, kg/m2, psi, mmH2O,
cmH2O, mH2O, Torr, atm, lb/ft2, inH2O, ftH2O, inH2O@4°C, ftH2O@4°C, mmHg,
cmHg, mHg, inHg, programmable.
Internals sensors
PLAGE
RESOL.
ACCURACY
-100 à 100 mbar
0,001 mbar
-500 à 500 mbar
0,01 mbar
-0,95 à 2 bar
0,01 mbar
0 à 2 bar (ABS)
0,01 mbar
-0,95 à 7 bar
0,1 mbar
-0,95 à 20 bar
0,1 mbar
0 à 20 bar (ABS)
0,1 mbar
Sur pression : 125 % F.S.
Manuel d'intructions NTA47248-300A4
ACCURACY
ACCURACY
Plus / XP models
BASIC model
±(0.01 % +0.1°C)
±(0.02 % +0.1°C)
±(0.01 % +0.2°F)
±(0.02 % +0.2°F)
±(0.01 % +0.1°C)
±(0.02 % +0.1°C)
±(0.01 % +0.2°F)
±(0.02 % +0.2°F)
±(0.01 % +0.1°C)
±(0.02 % +0.1°C)
±(0.01 % +0.2°F)
±(0.02 % +0.2°F)
±(0.01 % +0.1°C)
±(0.02 % +0.1°C)
±(0.01 % +0.2°F)
±(0.02 % +0.2°F)
±0,025 % F.S.
±0,025 % F.S.
±0,025 % F.S.
±0,025 % F.S.
±0,025 % F.S.
±0,025 % F.S.
±0,025 % F.S.
Port : 1/8 BSPF (femelle)
64