Press the [ ] key. Press the [ESC] key to return in measure mode without
changes.
Tc K
Tc J
Tc T
Tc R
Tc S
Pr e s s < > t o SET
Press the [ ] and [ ] keys to highlight the requested type of thermocouple.
Press the [ ] key to confirm the selection and going in measuring mode.
Press the [ESC] key to return in measure mode without changes.
5.4.2

External Cold Joint Reference Setting

Press the [MENU] key. The following page is shown:
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Pr e s s < > t o SET
Press the [ ] and [ ] keys to highlight the "Unit" option.
Press the [ ] key to confirm the selection. Press the [ESC] key to return in
measure mode without changes.
Un i t
Te mp e r a t u r e
Pr e s s u r e [ mb a r ]
Xs c a l e
Pr e s s < > t o SET
Select the "Temperature" option.
Press the [ ] and [ ] keys to highlight the "Rj Ch1 ext" option for channel 1
setting or highlight the "Rj Ch2 ext" option for channel 2 setting.
Press the [ ] key to confirm the selection.
Manuel d'intructions NTA47248-300A4
Rj i
MENU
Te mp e r a t u r e
Un i t
[ ° C]
Rj
Ch 1 e x t
[ 0 . 0 ]
Rj
Ch 2 e x t
[ 0 . 0 ]
Sc a l e [ I TS 9 0 ]
Pr e s s < > t o SET
Us e n u me r i c k e y s
[ ENTER]
Use the numeric pad (red keys) to insert the temperature value (e.g. 25.0°C)
Press [ENTER] to confirm or [] to go to the upper menu level without
changes.
Press the [ESC] key to return in measure mode without changes.
5.4.3

Temperature with RTDs

The calibrator can support 14 standard of resistance thermometers. See table on
Technical Specifications chapter.
Two, Three, or Four wire RTD connections can be performed for high accuracy
measurements. Refer to the electrical connection chapter.
Press the [SELECT] key until "SELECT CH1" or "SELECT CH2" is shown.
If you choose the Channel 1 setting the calibrator will display:
Pr e s s < > t o SET
If you choose the Channel 2 setting the calibrator will display:
N.B.
hannel 2 can measure only on Calys 120 IS model.
80
Rj
Ch 1 e x t
2 0 . 0
t o c o n f i r m
Rj
Ch 1 e x t
0 . 0
2 5 . 0
SEL ECT Ch 1
Th e r mo c o u p l e
RTD t h e r mo me t e r
Cu r r e n t
Vo l t a g e
Re s i s t a n c e