After installing the program, run it from "Programs" menu:
Connect the instrument to PC through the serial communication cable and
turn it on and enter the Firmware Upgrade mode by pressing [MENU]+[ON]
Select the right COM port (Serial port)
Select the same communication velocity (expressed in Baud) either on PC
(Baud rate) and on the instrument
ATTENTION
SET THE SAME BAUD RATE EITHER ON PC AND THE INSTRUMENT,
OTHERWISE NO COMMUNICATION HAPPENS.
IN CASE OF COMMUNICATION PROBLEMS, TRY TO DECREASE THE BAUD
Insert in the field "Filename" the position where the file with "HFW" extension
is, it means the firmware that has to be transferred to the instrument.
Press "Program" and wait the complete data transfer
Press "END" to quit the program
NOTE: eventually, press "Read me" to read an help online.
ATTENTION
Manuel d'intructions NTA47248-300A4
RATE.
IN CASE OF FAILED PROCEDURE, TURN THE INSTRUMENT OFF BY
PRESSING [ON/OFF] FOR ABOUT 10SECS AND REPEAT THE OPERATION.