IMPORTANT
BEFORE RUN THE PROCEDURE, BE SURE TO HAVE ALL MEMORIES WITH
THE SAME OUTPUT PARAMETERS. IF THE STORED MEMORIES HAVE
DIFFERENT PHYSICAL PARAMETERS (EG. MV AND MA) YOU COULD
SERIOUSLY DAMAGE THE LOAD.
•
Select the "On/Off" key to run the procedure. The display shown the following
page:
–
Ch 1
Ch 2 s c a n 0
1 0 0 . 0 2
The message "scan" is displayed in the status box when the sequencer is running. A
number shows the current step.
•
If the sequencer is programmed in "AUTO" mode, press the [HOLD] key to
pause and to continue the procedure.
•
If the sequencer is programmed in "MANUAL" mode, press the [HOLD] key to
go to next step.
5.12 Switch Test
Verifying functions are available for measurement's instruments of temperature or
pressure.
•
Press [MENU], the following page is shown:
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Pr e s s < > t o SET
Select "Advanced"
Manuel d'intructions NTA47248-300A4
1 5 . 2 0 0 0
mA
mV
MENU
Pr e s s < > t o SET
Select "Switch Test"
Pr e s s < > t o SET
You can also set the contact input for the switch test by selecting it between the
"external", "ch1" or "ch2" options:
Select "Switch..."
Pr e s s < > t o SET
and the select the contact input to set it
Ex t
Ch 1
Ch 2
Pr e s s < > t o SET
NOTE: the external contact input [EXT] is not available on IS models
If you have to verify the switch on a pressure, select "Pressure" and the type of the
sensor to use, internal or external:
94
Ad v a n c e d
Swi t c h Te s t
L e a k Te s t
HART
Tr x s i mu l a t o r
Gr a p h
Swi t c h Te s t
Pr e s s u r e
Te mp e r a t u r e
Swi t c h [ Ex t ]
Swi t c h Te s t
Pr e s s u r e
Te mp e r a t u r e
Swi t c h [ Ex t ]
Swi t c h