I n v a l u e
Ou t
Fu n c t i o n [ l i n e a r ]
I n e r r o r [ Er r o r ]
Sa mp l i n g t i me [ 1 ]
Pr e s s < > t o SET
•
On/Off: RUN and STOP the function
•
Source: Input channel (Ch1, P1, P2, PE, PA, RH%, T)
•
In value: Allows to program the low and full scale for the Trx source
•
Out value: Allows to program the low and full scale for the mA output (max.
range 0 to 50mA)
•
Function : Linearization mode (Linear, square, sqrt, log)
•
In error: Error message for input out of range (out=L/H, error)
•
Sampling time: It is possible to choose the interval between input data
acquisition (0.1, 1 or 10 seconds)
5.11 Memory Scan
The output channel (CH2) can be programmed for simulating a sequence of 10
values. This procedure uses the Channel 2 as output channel and the 1-9 memories.
Before run this feature, you have to set output values using consecutive memories
(see par. 4.3) and set channel 2 as output. .
Manuel d'intructions NTA47248-300A4
Tr x
v a l u e
•
Press the [MENU] key. The following page is shown:
Pr e s s < > t o SET
Select "Ramp-Memory scan"
Ra mp - Me mo r y s c a n
Pr e s s < > t o SET
Select the "Memory Scan"
Pr e s s < > t o SET
Set the parameters for simulation procedure
•
On/Off: runs and ends the function
•
Mode: AUTO: automatic sequence using the "Time" value as interval time –
MANUAL: require to press the [HOLD] key to go to the next step
•
From: First memory to be simulated.
•
To: Last memory to be simulated
•
Time: interval time between 2 steps
•
Mode: set the mode between the options "Auto" and "Manual"
93
MENU
Co n f i g u r a t i o n
Un i t
Ra mp - Me mo r y s c a n
Da t a l o g g e r
Ad v a n c e d
Ra mp
Me mo r y s c a n
Me mo r y s c a n
On / Of f
Fr o m [ 1 ]
To [ 8 ]
Ti me [ 2 ]
Mo d e [ AUTO]