connection not
including
Terminal A of
Channel 1,
Terminal A or
Terminal B of
Channel 2
Vo= 6.51 V
Io = 3 mA
Po= 19.7 mWatt
Co= 22 µF
Lo= 500 mH

14.2 ATEX Specifications

Model calys60/80/120 IS marked with label "Ex" has been built in accordance with
European Norm II 1G Ex ia IIC T4 and may be used in IEC defined zones 0,1 and 2.
The Intrinsically Safe (IS) is a security standard applied into potential explosive
environmental. The "Intrinsically Safe" certified instruments are been designed and
made in order to do NOT release thermal and electric energy able to induce the
ignition of flammable materials (gas) even in case of malfunction or connection
errors.
The calys60/80/120 IS are an ATEX directive compliant calibrator and it is certified
as II 1G Ex ia IIC T4
The explanation code is as follows:
II
Apparatus Grouping: all the applications except mines with Grisou.
1G
Category: permanent presence of explosive gas (zone 0) except zones with
explosion danger due to powders
Ex
Explosion protection by the European regulation
ia
Explosion protection type: the device or connector energetic level has been
limited unitl a secure level. By applying the Ui and Um maximum voltages
there are no risk to cause sparks in the following condition:
In presence of ALL the not-autorevealing failures with 1,5 of multiplying
coefficient for Ui and Um
In presence of ALL the not-autorevealing failures and one revealing failure
with 1,5 of multiplying coefficient for Ui and Um
In presence of ALL the not-autorevealing failures and two revealing
failures with 1 of multiplying coefficient for Ui and Um
IIC
Gas Grouping: compatible with the most dangerous gas; A=Propane,
B=Ethylene, C=Hydrogen
Manuel d'intructions NTA47248-300A4
T4
Temperature Class: maximum ambient operative temperature of 135°C.
The allowed ambient operating temperature is higher of the normal electrical
manufacture (from -20°C to +40°C), that's why there is the temperature
indication (-20°C Tamb +50°C)
The T4 temperature class has been adopted. All the temperature classes are
as follow:
T1 = 450°C of maximum superficial temperature
T2 = 300°C of maximum superficial temperature
T3 = 200°C of maximum superficial temperature
T4 = 135°C of maximum superficial temperature
T5 = 100°C of maximum superficial temperature
T6 = 85°C of maximum superficial temperature
X
The final "X" coefficient is for particular operating modes; in particular
calys60/80/120 IS are certified for low mechanical risk

14.3 Instructions for use

Special precautions for use in hazardous areas
1. Do not connect the output terminals to external powered circuits when
calibrator sources Volts or milliamps.
2. Make sure that the parameters of both the calibrator and the instrument
match.
3. Do not exceed parameters nameplate data.
4. Do not open battery enclosure in a potentially esplosive atmosphere.
5. Do not charge or attempt to charge the batteries.
6. Do not use the serial port.
7. Do not open the calibrator housing when an explosive gas atmosphere may
be present.