Once loaded data calibration, it is possible to check them selecting "Display
procedure". One or more pages appear, interchangeable by pressing
[ENTER]:
SEL ECT Pr o c e d u r e
Ta g [ Ta g 1 ]
Re s u l t s [ Fo u n d ]
Di s p l a y p r o c e d u r e
Di s p l a y r e s u l t s
Ru n p r o c e d u r e
Pr e s s [
]
t o SET
Ta g 1
Ac t u a l
l i mi t
± ( 0 . 0 2 5 % +2 . 0 mV)
Pr e s s [ ENTER]
Even from "Select procedure", it is possible to set the procedure as "As found" or
"As Left", it means to state whether the instrument is as found or as left after a
calibration
Fo u n d
L e f t
Pr e s s [
•
Now it is possible to run the procedure by pressing "Run procedure" and to
wait for its completion:
Manuel d'intructions NTA47248-300A4
Ta g 1
My Ta g 1
De s c r i p t i o n 1
De s c r i p t i o n 2
Re f .
L i mi t
± ( 0 . 0 1 % +2 . 0 mV)
Pr e s s [ ENTER]
Ta g 1
Te s t
Po i n t s :
0 . 0 0 0 0 |
3 . 0 0 0 0
1 . 0 0 0 0 |
4 . 0 0 0 0
2 . 0 0 0 0 |
5 . 0 0 0 0
2 . 5 0 0 0 |
Pr e s s [ ENTER]
Re s u l t
]
t o SET
OFF_ Vt o V
Re f : 5 . 0 0 0 0 V+0 . 0 1 2 5
Re f : Ou t
Ac t : I n
L i mi t :
Er r o r :
Pr o c e d u r e c o mp l e t e
Pr e s s [ ENTER]
When the Reference and/or the Actual values are out of limits, the correspondent
values will be in reverse mode.
Eventually, abort the procedure by pressing [ESC] key. A confirmation message will
be shown on display.
•
Finally, press [ENTER] as required at the end of the procedure. The calibrator
returns to the measuring mode.
•
Returning to procedure menu by pressing [CAL PROC] and selecting
"Display results", it is possible to check step by step the results of the
calibration (press [ENTER] to change the results pages).
SEL ECT Pr o c e d u r e
Ta g [ Ta g 1 ]
Re s u l t s [ Fo u n d ]
Di s p l a y p r o c e d u r e
Di s p l a y r e s u l t s
Ru n p r o c e d u r e
Pr e s s [
103
1 / 1
+5 . 0 0 0 0 V
+5 . 0 0 0 2 V
+0 . 0 0 3 5
+0 . 0 0 0 2
]
t o SET