Table 4. Enviromental conditions
Operating, ambient temperature
Operating, relative air humidity
Ambient pressure
Transport temperature
Transport relative humidity
Table 5. Transmitter function indicators light
Function indicators
Fast blinking green light
Green steady light
Alternating green/yellow blinking light
Yellow steady light
Yellow blinking light.
Table 6. Platforms compatible with PressureWire Aeris
PressureWire
Receiver
Ilumien
QUANTIEN
For other compatible platforms, check with your sales representative.
Compliance with Regulatory Requirements
Guidance and manufacturer's declaration – electromagnetic emissions
PressureWire system is intended for use in the electromagnetic environment specified below. The
customer or the user of PressureWire system should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
15°C to 35°C (59°F to 95°F) ) max ± 5°C temperature fluctuation
30-75%
425-850 mmHg
- 25°C to +60°C (-13°F to +140°F)
10-95%
Explanation
Zeroing (initializing)
Transmitter on, normal operation
Stability check failed (pulsatile pressure detected). Rezero
PressureWire at stable zero pressure.
PressureWire disconnected from transmitter. Reinsert
PressureWire firmly into transmitter.
Error. Possible causes can be that the battery level is too low for
operation or other internal error. Exchange PressureWire.
Compliance
Group 1
10
Electromagnetic environment –
guidance
The PressureWire system uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.