WARNING!
If dangerous gases were being pumped, flush the pump with air or inert gas
for at least 10 minutes prior to disassembly.
To test the Pumping System
1
2
3
Newly installed tip seals may require several hours of run time to seat properly
NOTE
and enable the pump to meet speed and base pressure specifications.
TPS-compact User Manual / 87-901-010-01
Run the TPS-compact for about 5 seconds. Verify that the front
fan is running. If you hear loud noises from the Scroll pump this
indicates that the tip seal or main O-ring are possibly out of
place.
Disassemble and repair as necessary.
The pump is now ready to return to service.
(C)
Technical Information
Service Options
253/262
17