17
Technical Information
RS 232/RS 485 Communication Description
The communication port mating connector is sup-plied with the
RS232 PCB (AMP/Cannon or equivalent 15-pin "D" type male
connector).
For example, the Transmit data signal from con-troller (pin 2) must
be connected to the host computer's receive data line (pin 2) and vice
versa. Consult the host computer's instruction manual for its serial
port connections.
Agilent cannot guarantee compliance with FCC regulations for radiated
NOTE
emissions unless all external wiring is shielded, with the shield being termi-
nated to the metal shround on the 0-subconnector. The cable should be secured
to the connector with screws.
Serial Communication Descriptions
This unit can communicate by two different proto-col:
These two protocols can be used as well with 232 or 485 media.
Please use "Window" protocol for new development.
NOTE
238/262
protocol (old system) "letter"
"Window" protocol (new system).
TPS-compact User Manual / 87-901-010-01
(C)