Service Options; Cleaning; Exterior; Interior - Agilent Technologies TPS-compact 969-8217 Notice De Mode D'emploi

Masquer les pouces Voir aussi pour TPS-compact 969-8217:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37

Service Options

Agilent pumping systems will provide many years of trouble-free
operation if the maintenance procedures are observed.
Pumping System cleaning and internal scroll pump tip seal
replacement are recommended when Pumping System TMP adsorbed
current (with blanked inlet flange) has risen to an unacceptably high
level (>500 mA) or when the message "TIP-SEAL SERV" appears on
the TPS-compact front side display.
The warning message appears after 5000 h of operation.
If your TPS-compact exhibits humming or grinding noises from the
scroll pump, a major overhaul should be done by Agilent Vacuum
Technologies. Advance exchange TPS-compact are available to
minimize downtime.
The parts needed for Tip-Seal replacement on the internal scroll
pump are available in the kit described in Orderable parts Table. This
kit contains seals and O-rings, and can be obtained from your Agilent
dealer.

Cleaning

Exterior

The exterior surfaces of the TPS-compact may be cleaned with mild
detergents only.

Interior

Internal dry scroll pump do not contain fluid for the cleansing of
accumulated dust and debris. Run the pump periodically at
atmosphere for a minute or two to flush it out.
TPS-compact User Manual / 87-901-010-01
(C)
Technical Information
Service Options
249/262
17

Publicité

Table des Matières
loading

Table des Matières