To Remove The Worn Tip-Seals - Agilent Technologies TPS-compact 969-8217 Notice De Mode D'emploi

Masquer les pouces Voir aussi pour TPS-compact 969-8217:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
Figure 42

To Remove the Worn Tip-Seals

Disconnect the TPS-compact from electrical power.
Unlock the scroll pump support by unscrewing the screws (A) (see
the "Tip-Seal replacement - Figure A")
1
2
3
4
5
TPS-compact User Manual / 87-901-010-01
Tip-Seal replacement - Figure C
Lift the TPS-compact in order to access the Scroll pump site
(you'll hear a "click" noise as soon as the upper part of TPS-
compact is locked).
Remove (4) M5 socket head bolts (B).
Remove the outboard housing axially off the frame (C).
Remove and discard the worn tip seals (D) and the main O-ring.
If compressed air is available, blow any remaining seal debris off
the scroll parts. If seal debris is attached to the sides, use a razor
or Exacto knife to scrape the debris off.