Emergency Stop

To immediately stop the TPS-compact in an emergency condition it is
necessary to remove the supply cable from the mains plug or moving
the external switch to OFF position.

Maintenance

The TPS-compact does not require any maintenance (except Tip-seal
replacement). Any work performed on the system must be carried
out by authorized personnel.
WARNING!
Before carrying out any work on the system, disconnect it from the mains,
vent the pump by opening the appropriate valve, wait until the rotor has
stopped turning and wait until the surface temperature of the pump falls
below 50 °C.
In the case of breakdown, contact your local Agilent service center.
Before returning the system to the constructor for repairs, the "Health and
NOTE
Safety" sheet attached to this instruction manual must be filled-in and sent to
the local sales office. A copy of the sheet must be inserted in the system
package before shipping.
If a system is to be scrapped, it must be disposed of in accordance
with the specific national standards.