Installation
Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
Do not install or use the pumping system in an environment exposed
to atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
In the presence of magnetic fields the pumping system must be
protected using a ferromagnetic shield.
The TPS-compact can be installed in any position. Fix the TPS-
compact in a stable position connecting the inlet flange of the
turbopump to a fixed counter-flange capable of withstanding a torque
of 50 Nm around its axis.
The turbopump with ISO inlet flange must be fixed to the vacuum
chamber by means of clamps or claws. The following table shows, for
each fixing device, the necessary number of clamps or claws and the
relevant fixing torque.
TPS-compact User Manual / 87-901-010-01
maximum pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C
relative humidity: 0 – 95 % (non-condensing)