10.3
Overview of process alarms (Selection)
Text from alarm
line
„DOOR State
Alarm"
„CO
Low Alarm"
2
„CO
High Alarm"
2
„Motor OVERLOAD
Alarm"
46
Meaning
Door alarm has triggered
The process value (CO
)
2
has fallen below its lower
alarm limit
- The process value (CO
)
2
has exceeded its upper
alarm limit
Motor protection cutout
has triggered
Display
Remedy
- Close door(s)
- Confirm alarm with
'ACK' (acknowledge)
- Check alarm upper
limits
- Check delay time,
extend if necessary
- Check CO2 supply
- Confirm alarm with
'ACK' (acknowledge)
- Check alarm upper
limits
- Check delay time and
extend if necessary
- Check CO
supply
2
- Confirm alarm with
'ACK' (acknowledge)
- Clear blockage in the
drive if possible
- Inform the service
department
- Confirm alarm with
'ACK' (acknowledge)