Stopping / cancelling a program

To stop or cancel a running program,
proceed as follows:
Step
Call up main "Batches"
menu.
Select mode.
Stop program.
Only a program stop/cancellation takes place.
All process values/states active at the time of the program stop/cancellation (temperature, speed, photo illuminance and
cooling) remain active.
The shaker continues to run with the values entered before the program stop/cancellation.
Changes are carried out directly in the respective controller for temperature (TEMP) or speed (SHAKER)
or at the photo illuminance (ILLUM) or cooling (COOL) switch.
- Carry out change of temperature or speed by pressing the "TEMP" or "SHAKER" key.
- Enter new value and confirm with "ok" or set controller mode to "off" and with confirm "ok".
- Make changes to the photo illuminance or cooling by pressing the "ILLUM" or "COOL" key.
- Switch photo illuminance or cooling off or on by pressing the "Off" or "On" key.
40
Key, Symbol, Display
Input
Press the "Batches" key.
Press the "Mode" key.
Press the "Stop" key.