10.4
Process messages
Text from alarm line
"Shut down" ... "Name"
"Name <3mA alarm ..."
"(Name of an analog input)
< 3 mA alarm"
10.5
Abbreviations
All designations of the functional elements in the
operating menus depend on the configuration of the
DCU system. Standard configurations use the
following designations:
10.5.1 Process values
Abbreviation
(TAG)
TEMP
SHAKER
10.5.2 Digital inputs
Abbreviation
(TAG)
TEMPC
HEATC
SHAKERC
FUSEC
DCUON
SUPON
48
Meaning
"EMERGENCY STOP" key was activated
Analog input with designated "name" has
defective cable
Meaning
Temperature on the sensor of the
heating
Shaker speed (speed during
shaking for CERTOMAT® systems)
Meaning (and measure if
necessary)
Warning message, check required
of the Pt 1000 in the air stream
Alarm signal, short circuit of the
heating
Alarm signal, short circuit of the
shaker drive
Alarm signal, safety fuse
actuated/replace fuse
Status signal from DCU main
switch - DCU switched on
Status signal from basic device -
basic device switched on
Remedy
Switch on device with the designated "name"
again (via EMERGENCY STOP key or main
switch)
Check if cable is trapped on peripheral devices
and check that it functions.
10.5.3 Digital outputs
Abbreviation
(TAG)
HEAT
TMPON
SHAON
Meaning (and measure if
necessary)
Status signal of the heating,
"Heating" in operation
Heating switched on
Shaker switched on