8.2.2.9
Alarms (IO-LINK events)
The electronics monitors the sensor and continuously compares the mechanical and ther
mal stresses against the limit values of the force transducer. With thermal monitoring,
they are also compared against the limit values of the electronic components.
The electronics uses a very high sample rate to evaluate the mechanical stress.
Even very short force peaks are recorded, and a notification is given if the limit values
are exceeded. As the output of measurement values via the IO-Link connection runs at
a lower sample rate, it may be that you cannot find a force value registered as a force
overshoot in the transferred measurement data.
60
Authoriza
Data type
tion
Read/
UInteger8T
Write
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
Data
Name
size
(bytes)
1
Nominal
Tempera
ture
Overload
Warning
4
Transducer
temperature
4
4
4
4
Hysteresis
for resetting
temperature
4
warnings
Description
Enables/
disables
warnings when
the tempera
ture exceeds/
falls below the
nominal (rated)
temperature of
the sensor.
0x00 = Disable
0x01 = Enable
Nominal (rated)
temperature
upper limit
Nominal (rated)
temperature
lower limit
Operating
temperature
upper limit
Operating
temperature
lower limit
Upper limits
Lower limits
U10M
ELECTRICAL CONNECTION