Index
Sub
(hex)
index
(hex)
0x0C46 0x00
0x0C47 0x00
0x0C26 0x00
Linearization via supporting points
u
Select "Stepwise linear Adjustment"; the "Adjustment supporting points" menu is dis
played. Open this menu.
u
Enter the number of supporting points, between 2 and 21. Note that the zero point is a
supporting point. So if you want to enter a straight line, select two supporting points.
("Adjustment Number of Supporting Points" menu item).
u
Under "Adjustment X" enter the force preset by the calibration system. Under "Adjust
ment Y" enter the measurement result shown in the calibration certificate that corre
sponds to the respective force.
u
It is important to start with the most negative force, as that is the highest tensile force.
On compressive force-only sensors , 0 N is defined as the "highest tensile force".
38
Authorizat
Data type
ion
ReadWrite
StringT
ReadWrite
StringT
ReadWrite
UIntegerT
Data
Name
size
(bytes)
32
Certificate
ID
32
Certificate
Expiration
Date
1
Lineariza
tion Mode
Description
Number of the
calibration
certificate
Date on which a
new calibration
is required
Selection of the
linearization
method:
0: No lineariza
tion is applied
1: Linearization
via supporting
points
2: Linearization
via cubic
function
U10M