Index
Sub
(hex)
index
(hex)
0x0DAD 0x00
0x0DAE 0x00
Tip
The digital switching outputs always work with the internal sample rate, and so are suitable
for very fast switching operations. The latency time between a physical event that triggers
a limit value switch in the amplifier module and a switchover of the digital switching output
is a maximum of 350 μs if no filters are used.
8.2.2.7.8 Statistical functions ("Statistics")
It is important to note that the internal sample rate is used to evaluate the signal in the
following functions. As the electronics works with 40,000 measurement points, even very
short load peaks are recorded. Note that any low-pass filters you set can quickly sup
press load peaks, which will then not be recorded in the maximum value memory.
All the following functions run continuously, and are not saved permanently, so a power
failure is equivalent to a reset.
50
Authoriz
Data type
ation
Read-
UIntegerT
Write
Read
UIntegerT
Write
Data
Name
size
(bytes)
1
Digital
Output
Pin
1
C/Q-Pin
function
in SIO-
Mode
Description
Select switching
channel to assign
to pin 2.
Permanent low
(0 V): 0x00
Permanent high
(24 V): 0x01
Switching
Channel 1: 0x02
Switching
Channel 2: 0x03
Select switching
channel to assign
to pin 4
Permanent low
(0 V): 0x00
Permanent high
(24 V): 0x01
Switching
Channel 1: 0x02
Switching
Channel 2: 0x03
U10M
ELECTRICAL CONNECTION