8.2.2.7.9 Reset functions
Four reset functions are available. All the reset functions are triggered by a correspond
ing system command (see "System command" section).
1. Device Reset
The sensor restarts. Note that the minimum and maximum values, as well as all of the
statistical information (Peak - Peak), will be lost. All the other settings and parameters
are retained.
2. Application Reset
The sensor does not restart. The following parameters are reset to their factory settings
or zero:
S
Filter settings
S
Switching points and hysteresis of limit value switches
S
Limit value switch teach function
S
Zero settings
S
Any selected conversions to another unit (measurement performed in N again)
S
Minimum and maximum values, as well as all of the statistical information
(Peak – Peak), will be lost.
S
Settings for digital inputs and outputs
S
Disabling of nominal overload warnings is activated