3. Restore Factory Reset
The sensor does not restart. In addition to the parameters specified in the application
reset, the entries in the "Application Tag", "Function Tag", and "Location Tag" fields are
reset.
Any linearization that may have been entered in the sensor (calibration certificate) is lost.
4. Back to box
All parameters are lost. Any overloads remain saved. The sensor resets to its delivery
condition. Any linearization that may have been entered in the sensor (calibration certifi
cate) is lost.
The system commands can be written directly to address "0x0002".
Index
Sub
(hex)
index
(hex)
0x0002 0
Code (decimal)
128
129
130
131
8.2.2.8
Additional information ("Diagnosis")
This menu item displays additional measured values and information.
Nominal Overload Warning: Here you can set whether the sensor is to generate an IO-
Link event on exceeding the nominal (rated) force ("Enable Warning") or not ("Disable
Warning"). Exceeding the maximum operating force always leads to an IO-Link event.
Nominal compressive force: Maximum nominal (rated) force in the compressive force
range.
Nominal tensile force: Maximum nominal (rated) force in the tensile force range.
For technical reasons, for compressive force transducers the same amount is entered as
for the maximum tensile force.
Operational compressive force: Maximum operating force in the compressive force
range.
Operational tensile force: Maximum operating force in the tensile force range.
Supply Voltage: Connected supply voltage.
54
Authorization
Write Only
Function
Device Reset
Application Reset
Restore factory settings
Back to box
Data
Data
type
size
(bytes)
UINT8
1
Description
System command
ELECTRICAL CONNECTION
U10M