Cleaning Cables
Lay the cable on a clean, flat surface. Position the connector ends of the cable to ensure that no liquid can enter the
inside of the connector during cleaning.
If at any time during the procedure the non-linting wipes become visibly soiled, discard soiled wipes and use more
clean wipes.
Use cotton swabs during the procedure, as needed, to clean and rinse seams, grooves, and narrow spaces.
To Clean
1.
Saturate a non-linting wipe with the validated cleaner.
2.
Saturate all surfaces of the product with the cleaner.
3.
Allow surfaces to remain visibly wet for 3-5 minutes.
4.
Dampen another non-linting wipe with cleaner.
5.
Wipe all surfaces of the product until they are visibly clean. Be sure to clean all sides of the product.
To Rinse
6.
Saturate a non-linting wipe with tap water.
7.
Start at the monitor-end connector and wipe all surfaces of the product at least one time.
8.
Visually inspect the product and repeat steps 6 and 7, if needed. If any part is not visibly clean, use a cotton swab
to clean between the grooves and narrow spaces.
9.
Using clean wipes, dry all surfaces of the product.
Disinfecting the Cable
Lay the cable on a clean, flat surface. Position the connector ends of the cable to ensure that no liquid can enter the
connector during disinfection.
If at any time during the procedure non-linting wipes become visibly soiled, discard soiled wipes and use more clean
wipes.
Use cotton swabs during the procedure, as needed, to disinfect and rinse seams, grooves, open areas, and narrow
spaces.
To Disinfect
1.
Saturate a non-linting wipe with the validated disinfectant.
2.
Thoroughly saturate all surfaces of the product.
3.
Allow surfaces to stay visibly wet for 10 minutes. Ensure that all seams, grooves and narrow spaces are saturated.
To Rinse and Complete Final Inspection
4.
Saturate a non-linting wipe with tap water.
5.
Start at the monitor-end connector and wipe all surfaces of the product.
6.
Repeat step 5 two more times, using more wipes as needed.
7.
Using clean wipes, dry the entire product.
8.
Visually inspect the product to ensure all surfaces are clean and dry.
Sterilizing
Never sterilize cables.