English
Instructions for Use
The cardiac output adapter cables are intended to connect catheters and injectate temperature probes to the cardiac
output input of Philips patient monitors. These products are for use only in combination with Philips patient
monitors when listed as accessories in the monitor's user documentation.
Recommended for
Right heart catheter measurement only
Transpulmonary measurement or Right heart catheter measurement
For warnings, cautions, intended use, operating instructions and specifications, see your patient monitor's
Instructions for Use.
Contraindications
The Cardiac Output Adapter Cable is contraindicated for use in an MRI environment.
Cleaning and Disinfection
Clean and disinfect Cardiac Output adapter cables before first use and before use on another patient.
Validated Cleaner and Disinfectant
Philips has validated the following cleaner and disinfectant which were tested for efficacy and use on the cables. These
agents were tested by Philips for both efficacy and prevention of damage to the Philips products being cleaned and
disinfected:
Cleaner:
•
Enzymatic detergent - composition:
Name
Alcohols, C9-11, ethoxylated surfactant
Glycerin
Citric Acid
N,N-Dimethyloctadecylamine oxide
Subtilisins (proteolytic enzymes)
•
Please refer to the detergent manufacturer's label for proper dilution and water temperature instructions.
Disinfectant:
•
70% Isopropyl Alcohol (Chemical Abstract Service #67-63-0)
Getting Started
Before cleaning and disinfecting the product:
1.
Visually inspect the cables for damage or deterioration. Check for any type of damage, such as tears, cracks, or
bent connector pins. Discard the product if damaged since there is a potential for incorrect readings.
2.
Always replace the product when there is extensive visible damage to the product (such as exposed wires or
breaks).
3.
Visually inspect the product for moderate to severe soiling. Products with moderate to severe soiling are not
intended to be cleaned or disinfected. If this situation occurs, dispose of the product.
4.
Secure the following supplies:
- non-linting wipes
- cotton swabs with tapered mini tips
- the validated cleaner and disinfectant using the concentration recommended by the manufacturer.
Length of monitor/
catheter cable
2.7 m/0.2 m
2.4 m/2.4 m
Percent %
1 - 5
1 - 5
1 - 5
1 - 2
0.1 - 1
Product Number
M1642A
M1643A