P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before performing preventive maintenance. At a minimum, check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more frequently based on your level of product usage.
N N o o t t e e - - Wash the exterior of the support surface before inspection, if applicable.
Inspect the following items:
Zipper and covers (top and bottom) are free of tears, cracks, cuts, holes, or other openings
Internal components for signs of staining from fluid ingress or contamination by fully unzipping the covers
Labels for legibility, proper adherence, and integrity
Foam and gel components have not degraded or come apart
Fire barrier for rips or other visible signs of damage (If excessive wear of the fire barrier is observed, replacement is
recommended)
Product serial number:
Completed by:
EN
10
Date:
1806-009-005 Rev AB.0