Télécharger Imprimer la page

Stryker TruRize Clinical Chair Mode D'emploi page 24

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 127
C C l l e e a a n n i i n n g g
C C A A U U T T I I O O N N
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until it is completely
dry and you have tested for safe operation.
To remove undesirable build-up prior to disinfecting between uses:
1. Hand wash all surfaces on the product with a mild detergent using spray or pre-soaked wipes.
2. Clean all exposed surfaces. Pay attention to high contact areas, such as the footrest cushion, seat cushion, backrest
cushion, footrest covers, armrest overmold and covers, and backrest mobility handles.
3. Follow the cleaning solution manufacturer's instructions for appropriate contact time and rinsing requirements.
4. Dry the product thoroughly before returning the product to service.
Avoid over saturation. Do not allow the product to remain wet.
EN
20
3333-009-005 Rev AA.0

Publicité

loading

Ce manuel est également adapté pour:

3333