N N o o t t e e : : U
is the a.c. mains voltage before applications of the test level.
T
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3333-009-005 Rev AA.0
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ' ' s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM radio bands
3 V/m
80 MHz to 2.7 GHz
3 Vrms
6 Vrms in ISM radio bands
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of T T r r u u R R i i z z e e ,
including cables, than the
recommended separation
distance calculated from the
equation appropriate for the
frequency of the transmitter.
Recommended separation
distance
D=(1.2) (√ P )
80 MHz to 800 MHz
D=(2.3) (√ P )
800 MHz to 2.7 GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
a
electromagnetic site
should be less than the
compliance level in each
frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
25
EN