Power save (sleep) mode activates after five minutes without button activation. Any contact on the operator control panel or
occupant control panel buttons will exit out of the power save (sleep) mode.
A A l l a a r r m m c c o o n n d d i i t t i i o o n n s s
There are three alarm conditions that may occur after the operator arms the chair exit system.
The chair exit alarm is an audible alarm (57 dB as measured per IEC 60601-1-8: 2012) that notifies the operator when any
of the following medium priority alarm conditions occur:
A A l l a a r r m m c c o o n n d d i i t t i i o o n n
Occupant exits the chair
Parking brake is unlocked with the chair exit system armed
Battery status indicator is low with the chair exit system
armed
N N o o t t e e - - The chair may activate the chair exit alarm if the chair cannot determine presence of an occupant. This may occur
when an occupant exits the chair long enough to activate the alarm, then sits back down.
To turn off the alarm, press the chair alarm arm/disarm on the operator control panel to disarm the chair exit system.
C C o o n n n n e e c c t t i i n n g g t t h h e e c c h h a a i i r r e e x x i i t t s s y y s s t t e e m m t t o o t t h h e e n n u u r r s s e e c c a a l l l l s s y y s s t t e e m m ( ( o o p p t t i i o o n n ) )
C C A A U U T T I I O O N N - - Always make sure that the nurse call system has been properly configured before use.
For communication between the chair exit system and your facility's nurse call system, connect the 1/4 in. nurse call
communication cable to the 1/4 in. phone connector on the head wall port.
This cable will default in a normally open configuration. If your nurse call system is setup as a normally closed configuration,
see the maintenance manual for instructions about how to change to a normally closed configuration. For chair exit system
alarm information, see Alarm conditions (page 16).
N N o o t t e e
• There will be a one second delay from when the occupant exits the chair to when the chair exit system notifies the
facility's nurse call system.
• The 1/4 in. nurse call communication cable limits are 0.5A, 42.4 VDC (30 VAC).
• The nurse call functionality in the medical device has not been evaluated for the requirements of Clause 17 (Normal
Operation) of UL 1069. The user is responsible to determine the operability of the chair exit system with all systems the
medical device is connected.
EN
16
R R e e s s u u l l t t
Occupant exited the chair and is at a potential risk of fall
Occupant is in an unsafe state to be left unattended and is
at a potential risk of fall. The chair stability has been
compromised as the chair may shift with occupant
movement.
Occupant is no longer being monitored by the chair exit
system and is at a potential risk of fall. The battery is at a
discharged state in which the chair will not be able to inform
the operator if the occupant exits the chair. If monitoring is
required, you must reconnect the chair to AC power.
3333-009-005 Rev AA.0