P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
Inspect the following items:
All welds
All fasteners are secure
Casters swivel and rotate
Casters are not loose or wobbly
Casters are free of wax and debris
Parking brake operates and all casters lock when you apply the parking brake
Backrest, seat, and footrest cushions are not ripped or torn
Armrests for cracks or splits
Armrests move, latch, and stow
Chair exit system alarms when armed while seat is unloaded and indicators operate
Chair exit system does not alarm when armed while the seat is loaded with more than 65 lb
All functions on operator control panels operate
All functions on occupant control panels operate
Batteries for replacement (every two years) (option)
Batteries can power product when AC is unplugged (option)
Batteries are charging (battery charging indicator) when you plug the power cord into a wall outlet (option)
1/4 in. nurse call communication cable functionality (option)
Power cord not frayed
Cables not worn or pinched
All visible electrical connections tight
All grounds secure to the frame
Ground impedance not more than 200 mΩ (milliohms)
Earth leakage current not more than 300 µA (microamps) for 120V or 500 µA (microamps) for 240V depending on
condition
Ground chain is clean, intact, and has at least two links touching the floor
Foley bag hook, armrest guide, and plastic covers are free from wear, tear, stresses. and mechanical damage
No rust or corrosion of parts