TROUBLESHOOTING
CODE
POSSIBLE CAUSE
Water or condensate in the sterilization
chamber.
A256
Air seepage through the gasket.
Vacuum pump failure.
Problem in the hydraulic circuit.
Water injection pump malfunction.
Problem in the hydraulic circuit.
A257
Steam generator safety thermostat triggered.
Steam generator malfunction.
Steam seepage through the gasket.
Excessive load.
A258
Problem in the hydraulic circuit.
Steam generator safety thermostat triggered.
Steam generator malfunction.
Drain filter obstructed.
A260
Problem in the hydraulic circuit.
Drain filter obstructed.
A261
Problem in the hydraulic circuit.
Drain filter obstructed.
A262
Problem in the hydraulic circuit.
Drain filter obstructed.
A353
Problem in the hydraulic circuit.
Drain filter obstructed.
A356
Problem in the hydraulic circuit.
Drain filter obstructed.
A360
Problem in the hydraulic circuit.
Drain filter obstructed.
A362
Problem in the hydraulic circuit.
ALARMS (CATEGORY A)
SUGGESTED SOLUTION
RESET according to the instructions.
Thoroughly dry the inside of the sterilization chamber and restart the
program.
DO NOT insert material impregnated with water or other liquids into the
chamber.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Contact Technical Service (see Appendix).
Contact Technical Service (see Appendix).
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water
and restart the program.
RESET according to the instructions.
Check that the load does not exceed the maximum values permitted.
(See the Summary Table in Appendix Technical Characteristics).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
Clean the chamber drain filter (see Maintenance Appendix).
Contact Technical Service (see Appendix).
87