ALARMS (CATEGORY A)
CODE
ALARM DESCRIPTION
A125
Faulty current draw detected
A126
Connection error with WiFi module
A131
Solenoid valve 1 failed
A132
Solenoid valve 2 failed
A133
Solenoid valve 3 failed
A134
Solenoid valve 4 failed
A135
Solenoid valve 5 failed
A136
Solenoid valve 6 failed
A137
Solenoid valve 7 failed
A140
Firmware update error
Pre-heating not executed within time-out
A201
(steam generator)
Pre-heating not executed within time-out
A202
(tube bundle heating element)
1st pulse with vacuum not achieved within
A250
time out
1st rise back up to atmospheric pressure not
A251
reached within the time-out
1st pressure pulse not reached within the
A252
time-out
2nd pulse with vacuum not achieved within
A253
time out
2nd rise back up to atmospheric pressure not
A254
reached within the time-out
2nd pressure pulse not reached within the
A255
time-out
3rd pulse with vacuum not achieved within
A256
time-out
1 = OK (warning)
2 = OK + door unlocking + RESET
(cont.)
MESSAGE ON THE DISPLAY
FAULTY
CURRENT DRAW
CONTACT TECH. SERVICE
WIFI MODULE ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 1 ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 2 ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 3 ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 4 ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 5 ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 6 ERROR
CONTACT TECH. SERVICE
SOLENOID VALVE 7 ERROR
CONTACT TECH. SERVICE
FW CLOUD UPDATE ERROR
STEAM GENERATOR
RESISTOR OPEN CIRCUIT
CONTACT TECH. SERVICE
HEATING BAND
OPEN CIRCUIT
CONTACT TECH. SERVICE
1PV TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
ATM1 UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
1PP UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
2PV TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
ATM2 UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
2PP UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
3PV TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
78
RESET MODE
2
1
2
2
2
2
2
2
1
2
2
2
2
2
2
2
2
2