16. APPENDIX – ALARMS
If the problem persists, contact the Technical Service (see APPENDIX) communicating the sterilizer model and serial number. This data is
indicated on the registration plate on the rear side of the device, on the declaration of conformity and can also be viewed by means of the
"sterilizer information" command.
Every time an anomalous condition occurs during the operation of the sterilizer, an alarm is generated, identified by a
specific code (consisting of a letter followed by a 3-digit number).
Alarm codes are divided into four categories:
E = ERROR/
WARNING
A = ALARM
H = HAZARD
S = SYSTEM
ERROR
In case of alarm, switch off the device only after having followed the indications displayed and having carried out the reset (see "Resetting the
system" paragraph).
16.1
ALARM INTERVENTION
The alarm intervention causes the cycle interruption (or the normal operation interruption), the display of the relevant
alarm code and message and an audible warning.
16.2 ALARM DURING A CYCLE
The alarm procedure is designed to ensure the user does not confuse an anomalous cycle with an efficiently carried out
one, and therefore eliminate the unintentional use of unsterilized materials; it is structured to guide the user to RESET the
sterilizer and follow the instructions below.
16.3 SYSTEM RESET
The system can be reset in two alternative ways, depending on the type of alarm (see the List of alarm codes below in
this appendix):
Pressing the OK button
Following the instructions displayed and holding down the RESET button for about 3 seconds
Press the RESET button for approx. 3 seconds to go back to
the main menu.
After the RESET and any technical operation necessary to
eliminate the fault, the device will be ready to perform a
new program.
Never turn off the device before carrying out the reset.
PROBLEM
SOLUTION
Incorrect handling and/or
The problem can normally
use or a cause outside the
be solved by the user.
device.
The problem can normally
First level fault
be solved on site by a
specialized technician.
The problem can normally
Second level fault
be solved by the Technical
Service Centre.
Electronic system error
(HW-FW).
CODE FORMAT
Exxx
(xxx = identification number 000 ÷ 999)
Axxx
(xxx = identification number 000 ÷ 999)
Hxxx
(xxx = identification number 000 ÷ 999)
Sxxx
(xxx = identification number 000 ÷ 999)
74