4.4
POWER SUPPLY
›
The electrical system to which the sterilizer will be connected must match the electrical characteristics of the
device.
›
Plate data are shown in the TECHNICAL CHARACTERISTICS table in the User Manual and on the back of the
machine.
4.5
ELECTRICAL CONNECTIONS
This information is shown on the back of the machine.
The sterilizer must be connected to a socket of the electric system having adequate capacity for the device and properly
grounded, in accordance with laws and/or regulations in force.
The manufacturer is not responsible for any damage caused by the installation of the sterilizer with unsuited and/or not properly
grounded electric systems.
Always connect the power cord directly to the power outlet. Do not use extensions, adapters or other accessories.
4.6
DIRECT CONNECTION TO A CENTRALIZED DRAINING POINT
›
Remove the cap holding clip and the cap on the rear of the autoclave.
›
Fit the plastic tube on the elbow union (supplied).
›
Fit the union and then refit the clip.
›
Fasten the clamp (supplied) to the drain siphon.
›
Cut the tube to the right length and insert its free end into the centralized draining point union locking it with the
dedicated ring nut.
Make sure that the tube is not bent, crushed or obstructed in any way.
The following diagram provides an arrangement of the components:
1.
At the centralized draining point
2.
Resting surface
3.
Clamp
4.
Drain siphon
The position of the union of the centralized draining point must be lower than the resting surface of the sterilizer. Otherwise, the tank may not be
emptied correctly.
If an automatic filling system (external pump or H2O AUX EV kit (AUX SV) is connected the use of the direct drain connection is required.
In case of fault or failure, this system allows any excess water produced by the automatic filling system to flow into the centralized draining point,
thus preventing flooding.
18