10.6
MANUAL INTERRUPTION
A cycle can be interrupted by the operator at any time, by
holding down the STOP button indicated in the figure
for about 3 seconds.
The command generates E999 error since the cycle could
not finish correctly.
Press ENTER to proceed.
If the cycle is interrupted during certain phases, an automatic cleaning procedure of the internal hydraulic circuit starts.
For the complete description of the alarms refer to "Alarms" appendix.
Press and hold RESET for about 3 seconds to open the
door.
After a manual interruption of the program, the load must not be used since the sterilization is not ensured.