14.2 SCHEDULED MAINTENANCE MESSAGES
The sterilizer periodically displays warning messages
relevant to "routine" maintenance operations that must be
carried out in order to ensure the proper operation of the
device.
Press DONE to confirm that the required maintenance
operation has been completed.
Press REMIND button to postpone the operation.
In this case, the warning message will reappear the next
time the sterilizer is used.
Always keep in mind the following general warnings:
›
DO NOT wash the sterilizer with direct jets of water, neither under pressure nor sprinkled. Seepage into electrical
and electronic components could irreparably damage the functioning of the device or its internal parts.
›
DO NOT use abrasive cloths, metal brushes (or other aggressive materials) or products for metal cleaning,
both solid and liquid, to clean the device and the sterilization chamber.
›
DO NOT use unsuitable chemical products or disinfectants to clean the sterilization chamber. In fact, these
products can cause irreparable damage.
›
DO NOT allow limescale or residues of other substances to accumulate in the sterilization chamber, on the
door and on the gasket, and remove these residues periodically. In fact, such residues may damage these parts,
and compromise the operation of the hydraulic circuit components.
The formation of white spots on the base of the internal wall of the chamber means that you are using poor quality demineralized water.
Before performing regular maintenance, make sure that the power cord plug is removed from the mains
socket.
If this is not possible, move the external switch of the device's power supply line to Off.
If the external switch is distant or not visible to the maintainer, place a "work in progress" sign on the switch,
after turning it off.
66