ALARMS (CATEGORY A)
CODE
ALARM DESCRIPTION
3rd rise back up to atmospheric pressure not
A257
reached within the time-out
3rd pressure pulse not reached within the
A258
time out
Chamber depressurization not reached within
A260
time out
Chamber levelling not reached within time
A261
out
Vacuum pulsation during drying not executed
A262
within time-out
1st drop to atmospheric pressure not
A353
completed within the time-out
2nd drop to atmospheric pressure not
A356
completed within the time-out
Vacuum pulsation after maintenance step not
A360
executed within time-out
Chamber depressurization during drying not
A362
reached within time-out
1 = OK (warning)
2 = OK + door unlocking + RESET
(cont.)
MESSAGE ON THE DISPLAY
ATM3 UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
3PP UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
ATM3 DOWNSTROKE TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
PRESSURE LEVELLING
TIMEOUT
CONTACT TECH. SERVICE
PD PRESSURE
UPSTROKE TIMEOUT
CONTACT TECH. SERVICE
ATM1 DOWNSTROKE TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
ATM2 DOWNSTROKE TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
SPD PRESSURE
DOWNSTROKE TIMEOUT
CONTACT TECH. SERVICE
PD PRESSURE
DOWNSTROKE TIMEOUT
CONTACT TECH. SERVICE
79
RESET MODE
2
2
2
2
2
2
2
2
2