17.3
HAZARDS (CATEGORY H)
CODE
ALARM DESCRIPTION
H150
MPX pressure sensor broken/not connected
H160
MPX pressure sensor short-circuited
Pconv/T ratio not balanced (Pconv>T)
H400
(STERILIZATION phase)
T/Pconv ratio not balanced (T>Pconv)
H401
(STERILIZATION phase)
Temperature over the MAX. limit
H402
(STERILIZATION phase)
Temperature below the MIN. limit
H403
(STERILIZATION phase)
Floating temperature over the limit
H404
(STERILIZATION phase)
Pressure over the MAX. limit (STERILIZATION
H405
phase)
Pressure below MIN. limit
H406
(STERILIZATION phase)
H410
Time measurement error
H411
Sterilization time error
Excessive pressure (sterilization chamber,
H990
MPX)
H991
Overheating (sterilization chamber, PT1)
H992
Overheating (steam generator, PT2)
H993
Overheating (layer resistance, PT3)
1 = OK (warning)
2 = OK + door unlocking + RESET
3 = Cycle failed + OK + door unlocking + RESET
MESSAGE ON THE DISPLAY
PRESSURE
SENSOR OPEN CIRCUIT
CONTACT TECH. SERVICE
PRESSURE
SENSOR SHORT-CIRCUIT
CONTACT TECH. SERVICE
INCORRECT P/T RATIO
CHECK LOAD
INCORRECT T/P RATIO
CHECK LOAD
TEMPERATURE BEYOND
MAXIMUM LIMIT
CONTACT TECH. SERVICE
TEMPERATURE BELOW
MINIMUM LIMIT
CONTACT TECH. SERVICE
ERRATIC TEMPERATURE
CONTACT TECH. SERVICE
PRESSURE BEYOND
MAXIMUM LIMIT
CONTACT TECH. SERVICE
PRESSURE BELOW
MINIMUM LIMIT
CONTACT TECH. SERVICE
INTERNAL TIMER ERROR
CONTACT TECH. SERVICE
STERILIZATION TIME
ERROR
PRESSURE BEYOND
MAXIMUM LIMIT
CONTACT TECH. SERVICE
PT1 OVERHEATING
CHECK LOAD
PT2 OVERHEATING
CONTACT TECH. SERVICE
PT3 OVERHEATING
CONTACT TECH. SERVICE
80
RESET MODE
3
3
2
2
2
2
2
2
2
2
2
2
2
2
2