–
TROUBLESHOOTING
CODE
POSSIBLE CAUSE
Door lock mechanism limit microswitch
failure.
E020
Door lock system gearmotor failure.
Door lock system lack of lubrication
Door lock mechanism limit microswitch
failure.
E021
Door lock system gearmotor failure.
E022
Door lock microswitches failure
E023
Door safety locking system failure.
Water level in the filling tank below minimum.
E030
MIN. water level sensor failure.
Water level in the drain tank over the MAX.
level.
E031
MAX. water level sensor failure.
Warning that the maximum water level in the
E042
tank has been reached (manual filling)
The autoclave cannot connect to LAN
E060
network
The autoclave cannot connect to WiFi
E061
network
E126
Cloud firmware is being updated
The cloud firmware version is not the correct
one with respect to the firmware process.
E141
There may be malfunctions in the connection
with WiFi / ethernet or cloud
Air seepage through the gasket.
E900
Problem in the hydraulic circuit.
Excessive humidity in the sterilization
chamber.
E901
Air seepage through the gasket.
Problem in the hydraulic circuit.
ERRORS (CATEGORY E)
SUGGESTED SOLUTION
RESET according to the instructions.
Contact Technical Service (see Appendix).
RESET according to the instructions.
Lubricate the door system
RESET according to the instructions.
Contact Technical Service (see Appendix).
RESET according to the instructions.
Contact Technical Service (see Appendix).
RESET according to the instructions.
Contact Technical Service (see Appendix).
RESET according to the instructions.
Top up the water to the MAX. level (or at least over the MIN. level).
Contact Technical Service (see Appendix).
RESET according to the instructions and empty the tank.
Completely drain the drain tank.
Contact Technical Service (see Appendix).
Interrupt the filling operation to prevent water spillage.
Make sure that configuration parameters of the Lan network are correct.
Check that the Lan network chosen for the connection is working properly.
Contact Technical Service (see Appendix).
Make sure that configuration parameters of the WiFi network are correct.
Check that the router managing the WiFi network is on and that the WiFi
network chosen for the connection is working properly.
Contact Technical Service (see Appendix).
Wait for the message to disappear and restart the machine
Contact Technical Service
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Contact Technical Service (see Appendix).
RESET according to the instructions.
Thoroughly dry the inside of the chamber and restart the program.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Contact Technical Service (see Appendix).
83
cont.