14. APPENDIX - MAINTENANCE
In addition to correct use, the user needs to perform regular maintenance in order to guarantee safe, efficient operation
over the device's entire life.
Always use personal protective equipment.
For better quality of maintenance, supplement routine checks with regular periodic check-ups that can be performed by
Technical Service Department (see Appendix).
It is also fundamental to perform a periodic sterilizer validation, i.e. a check of process thermo-dynamic parameters
and their comparison with the reference values detected by duly calibrated tools. Refer to 'Sterilizer Periodic Validation' in
the next part of the Appendix.
The regular maintenance described below consists in easy manual operations and preventative interventions involving
simple tools.
In the event of replacement of components or parts of the device, request and/or use original spare parts only.
14.1
REGULAR MAINTENANCE PROGRAMME
The table summarizes the maintenance interventions required to maintain the sterilizer in good working order.
In case of heavy use, we recommend shortening maintenance intervals:
DAILY
WEEKLY
MONTHLY
PERIODICALLY
A regular maintenance is essential to achieve the best performance of the device.
Periodically, a message will be displayed requesting that the above maintenance operations are performed.
For further information or in case of doubt, contact the Technical Service.
Clean the door gasket and the internal part of the door (14.3.1)
Clean external surfaces (14.3.3)
Clean the sterilization chamber and its accessories (14.3.2)
Clean the anti-dust filter (14.3.6)
Clean the internal water tank (14.3.9)
Warnings related to Scheduled Maintenance are displayed with the following frequency (14.2):
WARNING MESSAGE
CHAMBER FILTER CLEANING (14.3.4)
DOOR LOCK LUBRICATION (14.3.5)
ANTI-DUST FILTER CLEANING (14.3.6)
BACTERIOLOGICAL FILTER REPLACEMENT (14.3.7)
DOOR GASKET REPLACEMENT (14.3.8)
GENERAL SERVICE
65
FREQUENCY
250 CYCLES
or
3 MONTHS
or
250 CYCLES
3 MONTHS
or
500 CYCLES
6 MONTHS
500 CYCLES
or
6 MONTHS
or
1000 CYCLES
1 YEAR
or
3000 CYCLES