5.2
MAIN MENU
At the end of the starting procedure the main menu is
displayed on the side.
The sterilizer waits for the program selection (see "Program
selection" Chapter in the User Manual).
5.3
FILLING DEMINERALIZED / DISTILLED WATER
Only use demineralized / distilled water having the following characteristics to supply the sterilizer.
Description
DRY CONDENSATE
SILICONE OXIDE SiO2
IRON
CADMIUM
LEAD
HEAVY METAL RESIDUES (iron, cadmium
and lead excluded)
CHLORIDES
PHOSPHATES
CONDUCTIVITY AT 20° C / 68° F
pH VALUE
ASPECT
HARDNESS
MANUAL FILLING
1.
When filling the sterilizer for the first time during
installation, fill the reservoir until you hear the audible
beep and the LCD displays the message that Max Water
Level has been reached.
2.
After running cycles, if the LCD screen displays a Water
Level Low warning, fill the reservoir until you hear the
audible beep and the LCD displays the message that
Max Water Level has been reached.
Remove the cap from the top cover. Pour in water taking care
not to exceed the maximum level indicated inside the tank
(MAX.). Put the cap back in the top cover.
Pay attention not to spill water on the machine; if water is spilled, promptly dry.
The tank must be filled before the cycle starts or after its completion. Do not open the tank doors during the
cycle execution in order to prevent hot water leaks.
AUTOMATIC FILLING
For a listing of the automatic filling accessories available for the BRAVO G4, refer to the "ACCESSORIES" appendix in the
User Manual.
Values in the Water Supply
< 10 mg/L
< 1 mg/L
< 0.2 mg/L l
< 0.005 mg/L
< 0.05 mg/L
< 0. 1 mg/L
< 2 mg/L
< 0.5 mg/L
< 15 µS/cm
5 - 7
colourless, transparent, without
sediment
< 0.02 mmol/L
22
Values Inside Residual
< 1 mg/L
< 0. 1 mg/L
< 0. 1 mg/L
< 0.005 mg/L
< 0.05 mg/L
< 0. 1 mg/L
< 0. 1 mg/L
< 0. 1 mg/L
< 3 µS/cm
5 - 7
colourless, transparent, without
sediment
< 0.02 mmol/L